The bioactive Sclerostin ELISA kit is a 3.5 hour, 96-well sandwich ELISA for the quantitative determination of bioactive Sclerostin in human serum, EDTA-plasma, and citrate plasma. The human Sclerostin ELISA kit employs human serum-based standards to ensure the measurement of biologically reliable data.
The bioactive Sclerostin ELISA kit uses highly purified, epitope mapped antibodies with characterized binding kinetics.
Bioactive Sclerostin ELISA Assay Principle
The Biomedica Sclerostin ELISA is a sandwich enzyme immunoassay for the quantitative determination of bioactive Sclerostin (SOST) in human serum and plasma samples.
The figure below explains the principle of the bioactive Sclerostin sandwich ELISA:
In a first step, assay buffer is pipetted into the wells of the microtiter strips. Thereafter, standard/control/sample are pipetted into the wells, which are pre-coated with a recombinant monoclonal anti-human Sclerostin antibody. Any bioactive Sclerostin present in the standard/control/sample binds to the pre-coated antibody in the well. After incubation, a washing step is applied where all non-specific unbound material is removed. In a next step, the conjugate (anti-human Sclerostin-HRP) is pipetted into the wells and reacts with bioactive Sclerostin present in the sample, forming a sandwich. After another washing step, the substrate (tetramethylbenzidine; TMB) is pipetted into the wells. The enzyme-catalyzed color reaction of the substrate is directly proportional to the amount of bioactive sclerostin in the sample. This color change is detectable with a standard microtiter plate reader. A dose response curve of the absorbance (optical density, OD, at 450 nm) versus standard concentration is generated, using the values obtained from the standards. The concentration of bioactive Sclerostin in the sample is determined directly from the dose response curve.
Bioactive Sclerostin ELISA Typical Standard Curve
The figure below shows a typical standard curve for the bioactive human Sclerostin ELISA. The immunoassay is calibrated against recombinant human bioactive Sclerostin.
Bioactive Sclerostin ELISA Kit Components
Contents
Description
Quantity
PLATE
Recombinant monoclonal anti-human Sclerostin antibody pre-coated microtiter strips in a strip holder, packed in an aluminum bag with desiccant
Serum, EDTA plasma, citrate plasma, cell culture supernatant and urine are suitable for use in the human Sclerostin ELISA. Do not change sample type during studies. We recommend duplicate measurements for all samples, standards and controls. The sample collection and storage conditions listed are intended as general guidelines.
Serum & Plasma
Collect venous blood samples in standardized serum separator tubes (SST) or standardized blood collection tubes using EDTA or citrate as an anticoagulant. For serum samples, allow samples to clot for 30 minutes at room temperature. Perform separation by centrifugation according to the tube manufacturer’s instructions for use. Assay the acquired samples immediately or aliquot and store at -25°C or lower. Lipemic or haemolyzed samples may give erroneous results. Do not freeze-thaw samples more than five times.
Urine
Note: the experiments performed to measure bioactive Sclerostin in urine samples did not undergo a full validation according to FDA/ICH/EMEA guidelines. However, our performance check suggests that urine samples can be measured with this ELISA. For more information please refer to the validation data.
Aseptically collect the first urine of the day (mid-stream), voided directly into a sterile container. Centrifuge to remove particulate matter, assay immediately or aliquot and store at -25°C or lower.
Cell Culture Supernatant
Note: the experiments performed to measure bioactive Sclerostin in cell culture supernatant samples did not undergo a full validation according to FDA/ICH/EMEA guidelines. However, our performance check suggests that cell culture supernatant samples can be measured with this ELISA. For more information please refer to the validation data.
Remove particulates by centrifugation and assay immediately or aliquot and store samples at -25°C or lower. Do not freeze-thaw samples more than five times.
Reagent Preparation
Wash Buffer
1.
Bring the WASHBUF concentrate to room temperature. Crystals in the buffer concentrate will dissolve at room temperature.
2.
Dilute the WASHBUF concentrate 1:20, e.g. 50 ml WASHBUF + 950 ml distilled or deionized water. Only use diluted WASHBUF when performing the assay.
The diluted WASHBUF is stable up to one month at 4°C (2-8°C).
Standards for Serum & Plasma Measurements
1.
Pipette 250 µl of distilled or deionized water into each standard (STD) and control (CTRL) vial. The exact concentration is printed on the label of each vial.
2.
Leave at room temperature (18-26°C) for 15 min. Vortex gently.
Reconstituted STDs and CTRLs are stable at -25°C or lower until expiry date stated on the label. STDs and CTRLs are stable for at least four freeze-thaw cycles.
Standards for Cell Culture Supernatant Measurements
For the preparation of a cell culture-based standard curve always use the identical cell culture medium (CCM) as used for the experiment.
1.
Reconstitute standard 7 (STD7) in 250 µl deionized water. Leave at room temperature (18-26°C) for 15 min and mix well before making dilutions. Use polypropylene tubes.
2.
Mark tubes ccSTD6 to ccSTD1. Prepare a twofold serial dilution to obtain STD6 to STD2.
e.g. Dispense 100 µl cell culture medium into each vial.
3.
Pipette 100 µl of STD7 into tube marked as ccSTD6 (containing 100 µl cell culture medium).
Mix thoroughly.
4.
Transfer 100 µl of ccSTD6 into the tube marked as ccSTD5. Mix thoroughly.
5.
Continue in the same fashion to obtain ccSTD4 to ccSTD2. CCM serves as the ccSTD1 (0 nmol/l bioactive Sclerostin).
6.
Using the prepared standards, follow the protocol as indicated for serum samples.
Attention: Supplied STD1-STD7 and controls are only valid for serum and should not be used for cell culture measurements.
Sample Preparation
Bring samples to room temperature and mix samples gently to ensure the samples are homogenous. We recommend duplicate measurements for all samples.
Samples for which the OD value exceeds the highest point of the standard range can be diluted with ASYBUF (assay buffer).
Bioactive Sclerostin ELISA Assay Protocol
Read the entire protocol before beginning the assay.
1.
Bring samples and reagents to room temperature (18-26°C).
2.
Mark positions for STD/CTRL/SAMPLE (standard/control/sample) on the protocol sheet.
1.
Take microtiter strips out of the aluminum bag. Store unused strips with desiccant at 4°C in the aluminum bag. Strips are stable until expiry date stated on the label.
2.
Pipette 100 µl ASYBUF (assay buffer, red cap) into each well.
3.
Add 20 µl STD/CTRL/SAMPLE into the respective well. Swirl gently.
4.
Cover tightly and incubate for 2 hours at room temperature (18-26°C).
5.
Aspirate and wash wells 5x with 300 µl diluted WASHBUF (wash buffer). After final wash, remove remaining WASHBUF by strongly tapping plate against paper towel
6.
Add 100 µl CONJ (conjugate, amber cap) into each well. Swirl gently.
7.
Cover tightly and incubate for 1 hour at room temperature (18-26°C) in the dark.
8.
Aspirate and wash wells 5x with 300 µl diluted WASHBUF. After the final wash, remove remaining WASHBUF by strongly tapping plate against paper towel.
9.
Add 100 µl SUB (substrate, blue cap) into each well. Swirl gently
10.
Incubate for 30 min at room temperature (18-26°C) in the dark.
11.
Add 50 µl STOP (stop solution, white cap) into each well. Swirl gently
12.
Measure absorbance immediately at 450 nm with reference 630 nm, if available
Calculation of Results
Read the optical density (OD) of all wells on a plate reader using 450 nm wavelength (reference wavelength 630 nm). Construct a standard curve from the absorbance read-outs of the standards using commercially available software capable of generating a four-parameter logistic (4-PL) fit. Alternatively, plot the standards’ concentration on the x-axis against the mean absorbance for each standard on the y-axis and draw a best fit curve through the points on the graph. Curve fitting algorithms other than 4-PL have not been validated and will need to be evaluated by the user.
Obtain sample concentrations from the standard curve. If required, pmol/l can be converted into pg/ml by applying a conversion factor (1 pg/ml = 0.044 pmol/l (MW: 22.5 kDa)). Respective dilution factors must be considered when calculating the final concentration of the sample.
The quality control protocol supplied with the kit shows the results of the final release QC for each kit lot. Data for optical density obtained by customers may differ due to various influences and/or due to the normal decrease of signal intensity during shelf life. However, this does not affect validity of results as long as an OD of 1.50 or more is obtained for the standard with the highest concentration and the values of the CTRLs are in range (target ranges see labels).
Background & Therapeutic Areas
Sclerostin Protein
Sclerostin is a 22.5 kDa secreted glycoprotein that functions as a potent inhibitor of Wnt signaling. It acts by binding to the Wnt-coreceptor LRP5/6 thus inhibiting bone formation by regulating osteoblast function and promoting osteoblast apoptosis. The Sclerostin protein consists of two flexible N- and C-terminal arms and a cystine-knot with three loops, whereas the second loop binds to the LRP5/6 complex. Sclerostin is classically considered to be a monomeric protein, but data from Hernandez and colleagues (Hernandez et al., 2014) postulate that circulating sclerostin has a dimeric configuration. In addition, it is not yet well documented if also Sclerostin fragments circulate, but the comparison of different Sclerostin ELISAs suggest that fragments exist as well (Dallas et al., 2013).
Molecular weight
22.5 kDa
Cellular localization
Extracellular
Post-translational modifications
Glycosylation
Sequence similarities
Sequence similarity to the DAN (differential screening-selected gene aberrative in neuroblastoma) family of bone morphogenetic protein (BMP) antagonists
Sclerostin is nearly exclusively produced in osteocytes (van Bezooijen et al., 2009). Mutations in the Sclerostin (SOST) gene can cause sclerosteosis and van Buchem disease which are bone dysplasia disorders characterized by progressive skeletal overgrowth (Wergedal et al., 2003). Sclerostin levels are altered in response to hormonal stimuli or due to pathophysiological conditions. Sclerostin concentrations are increased in disorders such as hypoparathyroidism (Costa et al., 2011), Paget’s disease (Yavropoulou et al., 2012), multiple myeloma (Terpos et al., 2012) and in cancer induced bone diseases (Yavropoulou et al., 2012). Sclerostin levels are decreased in primary hyperparathyroidism (Lierop et al., 2010), as well as by the mechanical stimulation of bone (Robling et al., 2008). Several studies have found a positive association between sclerostin and bone mineral density (Amrein et al., 2012; Garnero et al., 2013). Sclerostin levels in chronic kidney disease (CKD) patients are increased up to 4-fold compared to patients without CKD and increase with CKD stage and declining kidney function (Cejka et al., 2012; Pelletier et al., 2013). In CKD patients, renal elimination of sclerostin increases with decreasing renal function (Cejka et al., 2014). In dialysis patients, sclerostin is an independent predictor of bone loss (Malluche et al., 2014). Numerous studies have shown that serum sclerostin levels are also associated with cardiovascular events (Kanbay et al., 2014; Viaene et al., 2013). The FDA authorization of a humanized monoclonal Sclerostin antibody for the treatment of osteoporosis in patients at high risk is currently under investigation (McClung, 2017). For reviews please refer to the following references: Costa et al., 2017; Drake and Khosla, 2017.
Bone Diseases
Osteoporosis
Cancer induced bone disease
Van Buchem disease
Thalassemia-associated osteoporosis
Cardiovascular Diseases
Vascular calcification in chronic kidney disease patients
Serum sclerostin and adverse outcomes in nondialyzed chronic kidney disease patients. Kanbay, M., Siriopol, D., Saglam, M., Kurt, Y.G., Gok, M., Cetinkaya, H., Karaman, M., Unal, H.U., Oguz, Y., Sari, S., Eyileten, T., Goldsmith, D., Vural, A., Veisa, G., Covic, A., Yilmaz, M.I., 2014. J. Clin. Endocrinol. Metab. 99, E1854-1861. PMID: 25057883
Sclerostin in mineralized matrices and van Buchem disease. van Bezooijen, R.L., Bronckers, A.L., Gortzak, R.A., Hogendoorn, P.C.W., van der Wee-Pals, L., Balemans, W., Oostenbroek, H.J., Van Hul, W., Hamersma, H., Dikkers, F.G., Hamdy, N. a. T., Papapoulos, S.E., Löwik, C.W.G.M., 2009. J. Dent. Res. 88, 569–574. PMID: 19587164
All Biomedica ELISAs are validated according to international FDA/ICH/EMEA guidelines. For more information about our validation guidelines, please refer to our quality page and published validation guidelines and literature.
Show validation literature
1. ICH Q2(R1) Validation of Analytical Procedures: Text and Methodology. 2. EMEA/CHMP/EWP/192217/2009 Guideline on bioanalytical method validation. 3. Bioanalytical Method Validation, Guidance for Industry, FDA, May 2018
Calibration
The bioactive Sclerostin immunoassay is calibrated against recombinant bioactive Sclerostin protein (Q9BQB4 (Uniprot ID)).
To determine the sensitivity of the bioactive Sclerostin ELISA, experiments measuring the lower limit of detection (LOD) and the lower limit of quantification (LLOQ) were conducted.
The LOD, also called the detection limit, is the lowest point at which a signal can be distinguished above the background signal, i.e. the signal that is measured in the absence of bioactive Sclerostin, with a confidence level of 99%. It is defined as the mean back calculated concentration of standard 1 (0 pmol/l of bioactive Sclerostin, five independent measurements) plus three times the standard deviation of the measurements.
The LLOQ, or sensitivity of an assay, is the lowest concentration at which an analyte can be accurately quantified. The criteria for accurate quantification at the LLOQ are an analyte recovery between 75 and 125% and a coefficient of variation (CV) of less than 25%. To determine the LLOQ, standard 2, i.e. the lowest standard containing bioactive Sclerostin, is diluted, measured five times and its concentration is back calculated. The lowest dilution, which meets both criteria, is reported as the LLOQ.
The following values were determined for the bioactive Sclerostin ELISA:
LOD
1.9 pmol/l
LLOQ
1.3 pmol/l
Bioactive Sclerostin ELISA Precision
The precision of an ELISA is defined as its ability to measure the same concentration consistently within the same experiments carried out by one operator (within-run precision or repeatability) and across several experiments using the same samples but conducted by several operators at different locations using different ELISA lots (in-between-run precision or reproducibility).
Within-Run Precision
Within-run (intra-assay) precision was assessed by measuring two samples of known concentrations three times within one bioactive Sclerostin ELISA lot by one operator.
ID
n
Mean bioactive Sclerostin [pmol/l]
SD [pmol/l]
CV (%)
Sample 1
3
19
0.3
1
Sample 2
3
153
1.0
1
In-Between-Run Precision
In-between-run precision (inter-assay) precision was assessed by measuring two samples seven times within two bioactive Sclerostin ELISA lots by two operators.
ID
n
Mean bioactive Sclerostin [pmol/l]
SD [pmol/l]
CV (%)
Sample 1
7
19
1.0
5
Sample 2
7
157
8.3
5
Bioactive Sclerostin ELISA Accuracy
The accuracy of an ELISA is defined as the precision with which it can recover samples of known concentrations.
The recovery of the bioactive Sclerostin ELISA was measured by adding recombinant bioactive Sclerostin to clinical samples containing a known concentration endogenous bioactive Sclerostin. The %recovery of the spiked concentration was calculated as the percentage of measured compared over the expected value. All our ELISAs are expected to have %recovery rates within 15% of the nominal value of the sample.
The table below shows the summary of the recovery experiments in the bioactive Sclerostin ELISA in different matrices:
% Recovery
+ 26 pmol/l
+ 110 pmol/l
Sample matrix
n
Mean
Range
Mean
Range
Serum
5
93%
76-111%
86%
82-95%
EDTA plasma
5
94%
85-104%
93%
86-98%
Citrate plasma
1
104%
-
99%
-
Show Individual Measurements
Data showing % recovery of recombinant bioacitve Sclerostin in human serum samples:
Bioactive Sclerostin [pmol/l]
Recovery (%)
Sample matrix
ID
Reference
+26 pmol/l
+110 pmol/l
+26 pmol/l
+110 pmol/l
Serum
s1
57
77
147
76
82
Serum
s2
75
104
180
111
95
Serum
s3
55
77
148
84
84
Serum
s4
44
67
140
90
87
Serum
s5
71
97
164
101
84
Mean
93
86
Data showing % recovery of recombinant bioacitve Sclerostin in human EDTA plasma samples:
Bioactive Sclerostin [pmol/l]
Recovery (%)
Sample matrix
ID
Reference
+26 pmol/l
+110 pmol/l
+26 pmol/l
+110 pmol/l
EDTA plasma
e1
190
212
295
85
96
EDTA plasma
e2
152
177
253
95
92
EDTA plasma
e3
165
190
259
97
86
EDTA plasma
e4
81
108
189
104
98
EDTA plasma
e5
67
91
171
91
94
Mean
94
93
Data showing % recovery of recombinant bioacitve Sclerostin in human citrate plasma samples:
Tests of dilution linearity and parallelism ensure that both endogenous and recombinant samples containing bioactive Sclerostin behave in a dose dependent manner and are not affected by matrix effects. Dilution linearity assesses the accuracy of measurements in diluted clinical samples spiked with known concentrations of recombinant analyte. By contrast, parallelism refers to dilution linearity in clinical samples and provides evidence that the endogenous analyte behaves in same way as the recombinant one. Dilution linearity and parallelism are assessed for each sample type and are considered acceptable if the results are within ± 20% of the expected concentration.
Dilution Linearity
Dilution linearity was assessed by serially diluting samples spiked with 110 pmol/l recombinant bioactive Sclerostin with assay buffer.
The table below shows the mean recovery and range of serially diluted recombinant bioactive Sclerostin in several sample matrices:
% Recovery of recombinant bioactive Sclerostin in diluted samples
1+1
1+3
1+7
Sample matrix
n
Mean
Range
Mean
Range
Mean
Range
Serum
6
98
93-103
86
73-100
89
75-103
EDTA plasma
6
102
97-106
99
97-103
91
91-109
Citrate plasma
1
119
-
132
-
103
-
Show Individual Measurements
Data showing dilution linearity of human serum samples containing recombinant bioactive Sclerostin:
Bioactive Sclerostin [pmol/l]
Recovery (%)
Sample matrix
ID
Reference
1+1
1+3
1+7
1+1
1+3
1+7
Serum
s1
147
71
37
17
96
100
93
Serum
s2
180
84
37
19
93
82
83
Serum
s3
259
134
62
27
103
96
82
Serum
s4
148
76
31
19
103
85
100
Serum
s5
140
67
27
18
96
78
103
Serum
s6
164
78
30
15
95
73
75
Mean
98
86
89
Min
93
73
75
Max
103
100
103
Data showing dilution linearity of human EDTA plasma samples containing recombinant bioactive Sclerostin:
Bioactive Sclerostin[pmol/l]
Recovery [%]
Sample matrix
ID
Reference
1+1
1+3
1+7
1+1
1+3
1+7
EDTA plasma
e1
295
151
72
35
102
97
96
EDTA plasma
e2
184
91
45
21
99
98
93
EDTA plasma
e3
253
128
62
29
101
97
91
EDTA plasma
e4
259
136
67
32
105
103
97
EDTA plasma
e5
189
101
46
22
106
97
95
EDTA plasma
e6
171
83
44
23
97
103
109
Mean
102
99
97
Min
97
97
91
Max
106
103
109
Data showing dilution linearity of human a citrate plasma sample containing recombinant bioactive Sclerostin:
Bioactive Sclerostin[pmol/l]
Recovery [%]
Sample matrix
ID
Reference
1+1
1+3
1+7
1+1
1+3
1+7
Citrate plasma
c1
174
103
58
22
119
132
103
Parallelism
Parallelism was assessed by serially diluting samples containing endogenous bioactive Sclerostin with assay buffer.
The table below shows the mean recovery and range of serially diluted endogenous bioactive Sclerostin in several sample matrices:
% Recovery of endogenous bioactive Sclerostin in diluted samples
1+1
1+3
1+7
Sample matrix
n
Mean
Range
Mean
Range
Mean
Range
Serum
7
100
89-108
103
96-108
106
90-120
EDTA plasma
6
105
99-111
108
99-125
123
107-154
Citrate plasma
2
91
89-94
91
86-96
103
102-104
Show Individual Measurements
Data showing dilution linearity of human serum samples containing endogenous bioactive Sclerostin:
Bioactive Sclerostin[pmol/l]
Recovery (%)
Sample matrix
ID
Reference
1+1
1+3
1+7
1+1
1+3
1+7
Serum
s1
139
67
33
17
97
96
95
Serum
s2
122
60
33
18
98
108
120
Serum
s3
114
61
29
13
108
101
90
Serum
s4
139
75
39
20
108
112
116
Serum
s5
103
50
28
15
98
108
116
Serum
s6
199
88
48
25
89
96
101
Serum
s7
89
46
23
12
104
105
106
Mean
100
103
106
Min
89
96
90
Max
108
108
120
Data showing dilution linearity of human EDTA plasma samples containing endogenous bioactive Sclerostin:
Bioactive Sclerostin[pmol/l]
Recovery (%)
Sample matrix
ID
Reference
1+1
1+3
1+7
1+1
1+3
1+7
EDTA plasma
e1
268
147
66
36
110
99
107
EDTA plasma
e2
210
109
52
30
104
99
115
EDTA plasma
e3
173
87
50
28
100
116
131
EDTA plasma
e4
184
98
47
28
106
102
122
EDTA plasma
e5
148
82
46
28
111
125
154
EDTA plasma
e6
242
120
66
32
99
110
107
Mean
105
108
123
Min
99
99
107
Max
111
125
154
Data showing dilution linearity of human citrate plasma samples containing endogenous bioactive Sclerostin:
Bioactive Sclerostin[pmol/l]
Recovery (%)
Sample matrix
ID
Reference
1+1
1+3
1+7
1+1
1+3
1+7
Citrate plasma
c1
175
82
42
23
94
96
104
Citrate plasma
c2
171
76
37
22
89
86
102
Mean
91
91
103
Bioactive Sclerostin ELISA Specificity
The specificity of an ELISA is defined as its ability to exclusively recognize the analyte of interest.
The specificity of the bioactive Sclerostin ELISA was shown by characterizing both the capture and the detection antibody through epitope mapping with overlapping peptides spotted to a microarray, characterization of binding kinetics with biolayer interferometry measurements and determination of antibody purity with size exclusion chromatography and affinity measurements. In addition, the specificity of the ELISA was established through competition experiments, which measure the ability of the antibodies to exclusively bind bioactive Sclerostin.
Epitope Mapping
Antibody binding sites were determined by epitope mapping using microarray analysis (Pepperprint GmbH).
Sclerostin Protein Structure - EPITOPES OF CAPTURE AND DETECTION ANTIBODY
Affinities of Capture and Detection Antibodies
Antibody affinities to bioactive Sclerostin were tested by biolayer interferometry measurements (Octet), which measures the binding of antibodies to a bioactive Sclerostin coated sensor.
The results of these measurements are shown in the figure below.
Both ELISA antibodies utilized in the bioactive Sclerostin ELISA bind to Sclerostin with high affinity. Biolayer interferometry measurements (Octet) of monoclonal capture antibody (mAb, pink) and polyclonal detection antibody (pAb, turquoise) binding to a sensor coated with sclerostin protein.
Both antibodies used in the bioactive Sclerostin ELISA bind to bioactive Sclerostin with high affinity.
Antibody Purity
Both the capture and detection antibody were purified by HPLC. The figures below reveal the high purity (>95%) of the antibody monomers used in the bioactive Sclerostin ELISA.
HPLC analysis of both antibodies. Size exclusion chromatography (SEC) of monoclonal capture antibody (mAb, pink) and polyclonal detection antibody (pAb, turquoise). The monoclonal capture antibody was analyzed using an Agilent Bio Sec column, whereas for the polyclonal detection antibody a Phenomenex Yarra X150 column was used.
Competition of Signal
Competition experiments were carried out by pre-incubating human samples containing endogenous concentrations of human bioactive Sclerostin with an excess of capture antibody (AB). The concentration measured in this mixture was then compared to a reference value, which was obtained from the same sample without the pre-incubation step. Mean competition was 100 % in all matrices tested.
Show Individual Measurement
Bioactive Sclerostin [pmol/l]
Sample matrix
ID
Reference
Reference + capture AB
% Competition
Serum
s1
28
28
100
Serum
s2
39
39
100
EDTA plasma
e1
40
40
100
EDTA plasma
e2
245
245
100
EDTA plasma
e3
88
88
100
Citrate plasma
c1
112
112
100
Citrate plasma
c2
108
108
99
Mean
100
Sample Stability
The stability of endogenous bioactive Sclerostin was tested by comparing measurements in samples that had undergone up to four freeze-thaw cycles.
For freeze-thaw experiments, samples were collected according to the supplier’s instruction using blood collection devices and stored at -80°C.
A set of samples (3 sera, 3 EDTA plasma, 2 citrate plasma) was aliquoted and freeze-thaw stressed. The reference samples are freeze thawed once. Samples can undergo four freeze-thaw cycles. The mean recovery of sample concentrations stressed by four freeze-thaw cycles is 93%.
Show Individual Mesurements
Bioactive Sclerostin concentrations of samples after freeze-thaw (F/T) cycles:
Bioactive Sclerostin [pmol/l]
Recovery (%)
4x F/T vs Reference
Sample matrix
ID
Reference
2x
3x
4x
Serum
s1
27
22
25
29
109
Serum
s2
40
35
33
35
87
Serum
s3
42
38
41
40
94
EDTA plasma
e1
236
224
231
226
96
EDTA plasma
e2
108
90
95
98
90
EDTA plasma
e3
98
85
91
91
93
Citrate plasma
c1
114
99
101
99
86
Citrate plasma
c2
114
108
100
100
88
Mean
93
Samples can undergo at least up to four freeze-thaw cycles.
Sample Values
Bioactive Sclerostin Values in Apparently Healthy Individuals
To provide expected values for circulating bioactive Sclerostin, a panel of samples from apparently healthy donors was tested.
A summary of the results is shown below:
Bioactive Sclerostin [pmol/l]
Sample matrix
n
Mean
Median
5% Percentile
95% Percentile
Min
Max
Serum
32
70.8
61.5
12.5
143.4
8
183
EDTA plasma
24
103.9
87
29.2
225.8
27
235
Citrate plasma
24
72.8
61.5
19.2
165.3
18
166
It is recommended to establish the normal range for each laboratory.
Plasma Bioactive Sclerostin Values in Kidney Transplant Recipients
Bioactive Sclerostin [pmol/l]
Sample matrix
n
Mean
Median
5% Percentile
95% Percentile
Min
Max
EDTA plasma
16
170.3
166.5
71
310
71
310
Serum Bioactive Sclerostin Values in a CKD Patient Cohort
Bioactive Sclerostin [pmol/l]
Sample matrix
n
Mean
Median
5% Percentile
95% Percentile
Min
Max
Serum
24
94.1
96
22.7
200.3
21
206
Matrix Comparision
To assess whether all tested matrices behave the same way in the bioactive Sclerostin ELISA, concentrations of bioactive Sclerostin were measured in serum, EDTA, and citrate plasma samples prepared from six apparently healthy donors. Each individual donated blood in all tested sample matrices.
A summary table of bioactive Sclerostin levels in various sample matrices is shown below:
Bioactive Sclerostin [pmol/l]
Donor ID
EDTA plasma
Citrate plasma
Serum
d1
89
74
57
d2
27
23
15
d3
65
43
39
d4
66
58
56
d5
66
57
56
d6
60
54
52
Measured values of human bioactive Sclerostin in serum are lower compared to plasma in an apparently healthy cohort (n=6).
It has been shown that Sclerostin values differ between serum and plasma even when these assays are validated in both matrices (McNulty et al., 2011). Measurements of Sclerostin in plasma are generally higher than in serum. The reasons for this difference are still unclear, however it is assumed that coagulation processes under conditions of serum collection might reduce the accessibility of recognizable determinants (Costa et al., 2017).
Figure showing matrix comparison of bioactive Sclerostin sample concentrations between serum, EDTA plasma and citrate plasma in an apparently healthy cohort (n=6).
Why is Heparın Plasma not Suggested as a Sample Matrix in this ELISA?
Heparın mainly binds to loop 2 and loop 3 of the Sclerostin molecule and disturbs the binding of the detection antibody utilized in this ELISA assay. For this reason, heparın plasma cannot be measured with this assay.
Comparison with other Assays
Biomedica’s Sclerostin ELISA (Cat.No. BI-20492*) was compared with the bioactive Sclerostin ELISA (Cat.No. BI-20472**). The same panel of samples was tested (16 EDTA plasma samples and 16 serum samples). The correlation between the two assays was R= 0.58.
*launched 2013, ** launched 2018
Show Individual Measurements
The correlation between the two assays resulted in R2=0.58. Sclerostin sample values measured with the Biomedica “bioactive Sclerostin ELISA” (Cat.No. BI-20472) are higher than in the Biomedica “Sclerostin ELISA” (Cat.No. BI-20492). The results demonstrate that the antibodies utilized in both assays bind to different regions of the Sclerostin molecule. The monoclonal capture antibody of the bioactive Sclerostin ELISA binds to the receptor binding site of Sclerostin, a region that is most probably more resistant to cleavage.